LAQUA PH1300

pH1300-S
pH mV(ORP) Ion
PCPRT USB
    Automatic pH calibration up to 5 points using USA, NIST, or Custom buffers
    Switchable pH resolutions - 0.1, 0.01, 0.001 pH
    Offset and slope display
    Ion calibration up to 5 points
    Automatic / Manual temperature compensation and temperature calibration function
    Auto-Hold / Auto-Stable measurement
    Auto log data; Built-in memory can store up to 999 data
    Real-time clock and date / time stamping
    Auto shut-off (programmable: 1 to 30 minutes)
    Calibration alarm (programmable: 1 to 400 days)
    Custom black and white static LCD
    RS232 and USB outputs for connecting to a GLP/GMP compatible printer or computer
    Security (password setting)
    2 years meter warranty

Meter Kit Available

The appearance of items in the picture are subject to change in each region.

Package Details:

    LAQUA PH1300 pH/ORP/Ion/Temp custom black and white static LCD meter
    Protection cover, integrated electrode stand, power adaptor with 6 plugs (US, EU, UK, ANZ, Korea, China)
    9615S-10D ToupH glass-body, refillable pH electrode with integrated temperature sensor
    101-SU: pH 4.01 / 7.00 / 10.01 buffers & 3.33 M KCl reference electrolyte
    Instruction manual

PH1300-SN: Kit with 101-S (pH 4.01 / 6.86 / 9.18 NIST buffers & 3.33M KCl) is available upon request.

Specifications

Model

LAQUA PH1300

pH/ORP/Ion/Temperature

pH Range

-2.000 to 20.000 pH

Resolution

0.1, 0.01, 0.001 pH

Accuracy

± 0.003 pH

Calibration Points

Up to 5

Buffer Options

USA, NIST, Custom

ORP Range

± 2000 mV

Resolution

0.1 mV

Accuracy

± 0.2 mV

Ion Range

0.000 µg/L to 9999 g/L

Resolution

4 significant digits

Accuracy

±0.3% F.S.

Calibration Points

Up to 5

Temperature Range

-30.0 to 130.0°C

Resolution

0.1°C

Accuracy

± 0.4°C

Calibration Option

Yes

Memory Data

999

Auto Log Data

Yes

Real-time Clock

Yes

Date / Time Stamp

Yes

Auto-Shut Off

Yes (Programmable: 1 to 30 minutes)

Auto-Hold / Auto-Stable

Yes

Offset / Slope Display

Yes (Independent acid and alkaline slopes depending on calibration)

Electrode Status

On screen display

Diagnostic Messages

Yes

Calibration Alarm

Yes (Programmable: 1 to 400 days)

Display

Custom black and white static LCD

Inputs

BNC, phono, DC sockets

Output

RS232, USB

Power Requirement

AC adaptor, 100-240V, 50/60Hz

Weight

500g

Dimensions

170(L) x 174(D) x 73(H) mm