The meters’ auto diagnostic function allows the user to make sure that the instruments will always give high performance and reliable measurements. Fully GLP/GMP compliant, the LAQUA can be connected to an external printer. To ensure traceability, all results are recorded together with sample name, user ID and calibration information.
We offer a comprehensive range of conductivity electrodes adapted to all type of application and sample. Recognised for their stability and longevity, HORIBA’s electrodes provide accurate measurement in all conditions.
LAQUA’s large capacitance touch screen panel:
The display is easy to understand and functions menu is shown
Switch between digital real time graphs or analogue displays during measurement with just flick of a finger
Full on-screen support on the meter via an electronic user guide with colour pictures that can be called-up at any time, a problem-solving guide for calibration and measurement
Functions
Easy navigation for meter and display available for electrode condition check
Full support for various pharmaceutical pure water guidelines (USP/EP/JP/CP)
Auto-hold function for calibration and measurement
Simultaneous connection to a GLP/GMP compatible printer and computer
Digital memory - maximum 2,000 sets of measurement data can be recorded
USB-PC communication and USB memory
Multi-language support (Japanese, English, Chinese, Korean)
Meter Kit Available
The appearance of items in the picture are subject to change in each region.
Package Details:
LAQUA DS-72 Cond/Res/Sal/TDS colour touch screen LCD meter
3552-10D Platinum/Platinum black, glass-body, k=1.0 conductivity electrode with integrated temperature sensor
103-S: 84 uS/cm, 1413 uS/cm, 12.88 mS/cm, and 111.8 mS/cm conductivity standard solutions
Electrode stand
Specifications
Model |
LAQUA DS-72 |
---|---|
Conductivity Range |
k=100 m-1:0.000 mS/m~19.99 S/m |
Method |
2 AC Bipolar |
Resolution |
0.05% of full scale |
Repeatability |
±0.5% ±1 digit |
Selectable Unit |
Yes |
Distilled Water Temperature Conversion |
Yes |
Periodical Check |
Yes |
JIS/CP/JP/USP/EP Pharmaceutical Water Application |
Yes |
Salinity Range |
0.00 to 80.00 ppt (0.000 to 8.000%) |
Method |
Conversion from conductivity |
Resolution |
0.01 ppt (0.001%) |
Salt Concentration Calibration |
Yes |
Resistivity Range |
k=100 m-1:0.00 Ω・m~199.9 kΩ・m |
Method |
Conversion from conductivity |
Resolution |
0.05% F.S. |
Repeatability |
±0.5%F.S. ±1 digit |
TDS Range |
0.01 mg/L to 1000 g/L |
Method |
Conversion from conductivity |
Resolution |
0.01 mg/L |
Temperature Range |
-30.0 to 130.0°C |
Resolution |
0.1°C |
Repeatability |
±0.1°C ±1 digit |
Input |
BNC, phono |
Output |
USB, RS232C, Analog |
Data Memory |
2000 |
Data Search |
Yes |
ID Input |
Yes |
Display |
Touch Screen Color LCD |
Languages |
English, Japanese, Chinese, Korean |
Navigation Function |
Yes |
Temperature Compensation |
Auto / Manual |
Auto-Hold Setting |
Yes |
Password Setting |
Yes |
Software Upgrade Function |
Yes |
Operating Temperature |
0 to 45°C |
Power Requirement |
AC adptor, 100-240V 50/60Hz |
Dimensions |
170(W) x 174(D) x 73(H) mm |
Weight |
700g |